Performance evaluation of five ELISAkits for detecting anti-SARS-COV-2 IgG antibodies
Objectives: To evaluate and compare the performances of five commercial ELISA assays (EDI, AnshLabs, Dia.Pro, NovaTec, and Lionex) for detecting anti-SARS-CoV-2 IgG.
Methods: 70 negative control samples (collected before the COVID-19 pandemic) and samples from 101 RT-PCR-confirmed SARS-CoV-2 patients (collected at different time points from symptoms onset: ≤7, 8-14, and >14 days) were used to compare the sensitivity, specificity, agreement, positive and negative predictive values of each assay with RT-PCR. A concordance assessment between the five assays was also conducted. Cross-reactivity with other HCoV, non-HCoV respiratory viruses, non-respiratory viruses, and nuclear antigens was investigated.
Results: Lionex showed the highest specificity (98.6%, 95%CI: 92.3-99.8), followed by EDI and Dia.Pro (97.1%, 95%CI: 90.2-99.2), NovaTec (85.7%, 95%CI: 75.7-92.1), then AnshLabs (75.7%, 95%CI: 64.5-84.2). All ELISA kits cross-reacted with one anti-MERS IgG positive sample except Lionex. The sensitivity was low during the early stages of the disease but improved over time. After 14 days from symptoms onset, Lionex and NovaTec showed the highest sensitivity at 87.9% (95%CI: 72.7-95.2) and 86.4% (95%CI: 78.5-91.7), respectively. The agreement with RT-PCR results based on Cohen’s kappa was as follows: Lionex (0.89)> NovaTec (0.70)> Dia.Pro (0.69)> AnshLabs (0.63)> EDI (0.55).
Description: Blocking Buffer for antigen-down and sandwich ELISA. PanBlock is for antigen-down and sandwich ELISAs and for those with high background problems or cross-react with mammalian buffers.
Description: Blocking Buffer for antigen-down and sandwich ELISA. PanBlock is for antigen-down and sandwich ELISAs and for those with high background problems or cross-react with mammalian buffers.
Simultaneous quantification of rituximab and eculizumab in human plasma by liquid chromatography-tandem mass spectrometry and comparison with rituximab ELISAkits
Objectives: Specific and sensitive analytical techniques to quantify therapeutic monoclonal antibodies (mAbs) are required for therapeutic drug monitoring. The quantification of mAbs has been historically performed using enzyme-linked immunosorbent assays (ELISAs), for which the limitations in terms of specificity have led to the development of alternative analytical strategies.
Methods: Here, we describe the validation of a liquid chromatography tandem mass-spectrometry (LC-MS/MS) method for the simultaneous quantification of rituximab (RTX – anti-CD20) and eculizumab (ECU – anti-C5). Sample preparation was based on our previously published method, using protein G purification and trypsin digestion.
A new specific peptide for RTX, containing an N-terminal pyroglutamine and a trypsin miss-cleavage, enables better sensitivity, while peptide of ECU was chosen thanks to an in silico trypsin digestion and the Skyline® software. Full-length stable-isotope-labeled adalimumab was added to plasma samples as an internal standard. RTX in 50 human serum samples was quantified by LC-MS/MS and the concentrations obtained compared to those obtained with two commercial ELISA kits (Lisa Tracker® and Promonitor®).
Results: Calibration curves were linear from 1 to 200 µg.mL-1 for RTX and 5 to 200 µg.mL-1 for ECU, and within-day and between-day accuracy and precision fulfilled Food and Drug Administration validation criteria. Comparison of the LC-MS/MS method with ELISA showed a negligible bias with the Lisa Tracker® kit (4%), but significant bias with the Promonitor® assay (mean underestimation of 69% for the Promonitor® assay).
Comparison of an Immunochromatographic Assay Kit with DAS-ELISA for Large-Scale Diagnosis and Molecular Discrimination of Satsuma Dwarf Virus Collected from Citrus Orchards
Satsuma dwarf virus (SDV) seriously damages citrus production by reducing the quality and yield of fruit. To avoid contamination with SDV, mother trees are checked to be SDV-free in advance of nursery tree distribution. In this study, we compared an
immunochromatographic assay (ICA) kit with double-antibody sandwich enzyme-linked immunosorbent assay (DASELISA) for large-scale diagnosis of SDV in orchardgrown trees in Shizuoka Prefecture, Japan. The two methods gave conflicting results for 11 of 1,705 samples, all of which were negative by DAS-ELISA but positive by ICA.
The samples scored as positive by either DASELISA or ICA were analyzed by reverse transcription polymerase chain reaction and all were confirmed to be positive. These results validate the use of ICA as a screening method for large-scale diagnosis. Strain discrimination revealed that 16 of 22 isolates belonged to SDV, while citrus mosaic virus (CiMV) infection only and co-infection (SDV and CiMV) were in a minority.
How Reliable are Commercially Available Glypican4 ELISAKits?
Objective: Glypican4 is an interesting new adipokine, which seems to play an important role in developmental processes and is potentially associated with metabolic changes in obesity and type 2 diabetes mellitus. Currently, only a few studies examined glypican4 in human blood, mainly in adults.
Design, patients and measurements: The aim of our study was to investigate glypican4 serum levels in lean, overweight, and obese children and adolescents, to unravel a possible association between glypican4 serum levels and parameters of obesity and insulin resistance. In order to determine a suitable method for investigating glypican4 serum levels, we validated two commercially available human glypican4 ELISA kits, using serum and plasma samples of an obese, insulin-resistant patient, and a healthy control subject, a human recombinant glypican4 protein fragment and glypican4-overexpressing cell lysate.
Results: Using ELISA kit #1 we were not able to detect values above background level, apart from standard curve values. ELISA kit #2 initially seemed suitable to measure glypican4, but further validation experiments showed non-linearity of serial dilutions, no recognition of a human recombinant glypican4 protein fragment and non-linearity in the recovery of glypican4-overexpressing cell lysate. In addition, there was a considerable decrease (approx. 68%) of measured values between two experiments, performed at different time points with aliquots of the same serum sample. Contrary to that, further experiments found sample stability not to be compromised.
Evaluation of three commercially available ELISAkits for the determination of chromogranin A
Chromogranin A (CgA) is currently the most valuable tumor biomarker for diagnostic work-up, management and follow-up of neuro-endocrine tumors (NET). The aim of our study was to compare three different commercially available CgA ELISA kits and to evaluate their analytical and clinical performance.
CgA was measured with three different commercial ELISA kits on leftover sera from 40 patients: Chromoa R assay (Cis), Hu chromogranin A ELISA (Dia), and Neolisa chromogranin A (Euro). Analytical and clinical performance was evaluated by measuring precision, area under the ROC curve, sensitivity, specificity, positive and negative predictive value, correlation coefficients and Passing and Bablok regression analyses.
Precision (CV%) was acceptable for all evaluated ELISA’s (Cis 10.3%; Dia 9.8%; Euro 14.5%). The area under the curve (AUC) was comparable between the three assays (Cis 0.693; Dia 0.627; Euro 0.721). Sensitivity varied between 41.2% and 64.7%. Specificity ranged between 69.6% and 82.6%. Pairwise comparison revealed significant systematic and proportional differences when comparing Cis versus Dia (Cis = 25.30 + 1.94 Dia) and Euro versus Dia (Euro = 26.54 + 1.92 Dia).
Analytical and clinical performance was comparable for the three ELISA’s. CgA results obtained with different ELISA’s are not interchangeable. Abbreviations CgA: Chromogranin A; ELISA: Enzyme-linked immunosorbent assay; IRMA: Immunoradiometric assay; NET: Neuroendocrine tumors; PPI: Proton-pump inhibitors; RIA: Radioimmunoassay.
Inter-laboratory comparison of 2 ELISAkits used for foot-and-mouth disease virus nonstructural protein serology
Serologic assays used to detect antibodies to nonstructural proteins (NSPs) of foot-and-mouth disease virus (FMDV) are used for disease surveillance in endemic countries, and are essential to providing evidence for freedom of the disease with or without vaccination and to recover the free status of a country after an outbreak. In a 5-site inter-laboratory study, we compared the performance of 2 commercial NSP ELISA kits (ID Screen FMD NSP ELISA single day [short] and overnight protocols, ID.Vet; PrioCHECK FMDV NS antibody ELISA, Thermo Fisher Scientific).
The overall concordance between the PrioCHECK and ID Screen test was 93.8% (95% CI: 92.0-95.2%) and 94.8% (95% CI: 93.1-96.1%) for the overnight and short ID Screen incubation protocols, respectively. Our results indicate that the assays (including the 2 different formats of the ID Screen test) can be used interchangeably in post-outbreak serosurveillance.